Other gas sampling pumps distributed by Nextteq®

Nextteq® International LLC is also distributes the Gastec-brand gas sampling pump.  Nextteq® continues to supply custom inventory management and technical support for any Nextteq® Distributor who carries this product.


Configurations Available

Gastec GV-100  gas sampling pump kit.

Video Instructions

Operating Procedures

The Gastec® GV-100 Pump operating procedure : 1. Break off both ends of the detector tube. Insert the tube into the rubber inlet.

1. Break off both ends of the detector tube. Insert the tube into the rubber inlet.

The Gastec® GV-100 Pump operating procedure : 2. Pull out the handle until it is locked. Wait until the sampling time has elapsed.

2. Pull out the handle until it is locked. Wait until the sampling time has elapsed.

The Gastec® GV-100 Pump operating procedure : 3. Read the measurement at the end of the coloured layer.

3. Read the measurement at the end of the coloured layer.

Airtightness Test

If the sampling pump is not airtight, measurements will not yield correct results. Before starting to measure, be sure to verify that the sampling pump is really airtight.

The Gastec® GV-100 Pump airtightness test : 1. First, confirm that the inlet nut is not loose.

1. First, confirm that the inlet nut is not loose.

The Gastec® GV-100 Pump airtightness test : 2. Insert a detector tube (with both tips intact) into the rubber tube inlet of the sampling pump.

2. Insert a detector tube (with both tips intact) into the rubber tube inlet of the sampling pump.

The Gastec® GV-100 Pump airtightness test : 3. With the handle fully pushed in, align both of the guide marks (red line + triangle) on the pump (tail-end block) and the handle (▲100).

3. With the handle fully pushed in, align both of the guide marks (red line + triangle) on the pump (tail-end block) and the handle (â–²100).

The Gastec® GV-100 Pump airtightness test : 4. Pull out the handle in one thrust until it is locked, and then release the handle. Wait until the sampling time (approx. 1 minute) has elapsed.

4. Pull out the handle in one thrust until it is locked, and then release the handle. Wait until the sampling time (approx. 1 minute) has elapsed.

The Gastec® GV-100 Pump airtightness test : 5. Grasp the handle and turn it 90 degrees. If the pump is completely airtight, the handle should readily return to its original position; with the handle's demarcation read line fully concealed.

5. Grasp the handle and turn it 90 degrees. If the pump is completely airtight, the handle should readily return to its original position; with the handle's demarcation read line fully concealed.


Note:

When you return the piston, please ease it back gradually (lightly holding on to it so it does not snap back in one thrust. Since there is a vacuum inside the cylinder, the handle tends to return rapidly). If it snaps back too hard, the impact can damage the pump causing it to malfunction.